Important Information for G5 AED Owners & Users
⚠️ URGENT: ZOLL Powerheart® G5 AED Correction – Action Required
ZOLL has issued an urgent correction for Powerheart® G5 AEDs due to a self-test error that could prevent the device from delivering a shock. The issue is linked to humidity exposure beyond the device’s specifications. Check your AED immediately and ensure it is stored within proper conditions.
Steps to follow:
✅ Monitor the status of the device and self-tests periodically.
✅ Store AEDs in proper temperature & humidity
✅ Ensure a program is in place that periodically monitors the status of these devices.
✅ Confirm receipt of this notice with ZOLL
✅ Monitor the status of the device and self-tests periodically.
✅ Store AEDs in proper temperature & humidity
✅ Ensure a program is in place that periodically monitors the status of these devices.
✅ Confirm receipt of this notice with ZOLL
👉 For those who experience any self-test failure, take action to resolve any alerts as soon as possible. You may contact ZOLL’s technical support to troubleshoot the AED.